    Split

    Split injections are ideal for samples with high concentrations, high matrix loads, or when the limits of detection do not need to be low. The reason for that is that during injection, the split vent is open and only a small part of the total flow -and therefore of the sample and matrix- is transferred into the column, avoiding column overload, and prolonging its lifetime. These liners can also be used for other applications, such as the analysis of polymers in combination with a pyrolyzer.

    Standard recommendation:

    227-35007-01 - for GC-2030
    221-75193      - for GC-2010
    221-75195      - for other models

     

    Specifications & Ordering Information

    P/N   Wool Description
    227-35007-01
    227-35007-01
    Yes Standard Split Liner for GC-2030
    221-75194
    221-75194
    No Standard Split Liner for GC-2030 and GC-2010
    221-75195
    221-75195
    Yes Standard Split Liner for GC-2025 and GC-2014
    221-41444-05
    221-41444-05
    No Standard Split Liner for GC-2025 and GC-2014
    221-75193
    221-75193
    Yes Standard Split Liner for GC-2010

     

    Each P/N includes 5 deactivated liners. (Standard deactivation: max. temp. 400 °C)

    All liners presented here are compatible with all Shimadzu SPL-models. We recommend using the corresponding liner for each injector, as the position of the wool - or the indentations where the wool would be - has been specifically designed for each injector.

     Special deactivations

    Specifications & Ordering Information

    P/N   Wool/Frit Description
    961-01479-05
    961-01479-05
    Wool

    SPL Liner for basic compounds (e.g. amines);

     base deactivated
    961-01479-06
    961-01479-06
    CarboFrit
    961-01479-08
    961-01479-08
    Wool

     SPL Liner for reactive compounds (e.g. chlorinated pesticides);

     Siltek deactivated
    961-01479-09
    961-01479-09
    CarboFrit

     

    Each P/N includes 5 deactivated liners. Suitable for all injector models. The maximum temperatures depend on the deactivation:

    -    Standard deactivation: 400  °C
    -    Base deactivation: 350  °C
    -    Siltek deactivation: 380 °C

     

    Splitless

    Splitless injections are ideal for samples with low concentrations, or when the limits of detection need to be low (e.g., trace analysis). The reason for that is that at the beginning of the injection, the split vent is closed, and all the flow is directed into the column. After a set amount of time, the vent is opened, and any remaining solvent is flushed out. 

    Standard recommendation:

    227-35008-01 - for standard applications
     

    Specifications & Ordering Information

    P/N   Wool Description
    227-35008-01
    227-35008-01
    Yes Standard Splitless Liner; standard deactivated
    221-75192
    221-75192
    No Splitless Liner; standard deactivated
    961-01480-08
    961-01480-08
    Yes

    Splitless Liner for reactive compounds (e.g. chlorinated pesticides);

    Siltek deactivated
    980-14622
    980-14622
    Yes

    Splitless Liner for Quick-DB applications (e.g. pesticides);

    Sky deactivated

     

    Each P/N includes 5 deactivated liners. Suitable for all injector models. The maximum temperatures depend on the deactivation:

    -    Standard deactivation: 400 °C
    -    Siltek and Sky deactivation: 380 °C

     

    Split / Splitless

    Sometimes an instrument is used for split and splitless applications, and changing the liners accordingly is not an option. In those cases, split/splitless liners are recommended.

    Standard recommendations:

    227-35007-01 - for GC-2030
    221-75193      - for GC-2010
    221-75195      - for other models
     

    Specifications & Ordering Information

    P/N   Wool Description
    227-35007-01
    227-35007-01
    Yes Standard Split/Splitless Liner for GC-2030
    221-75194
    221-75194
    No Standard Split/Splitless Liner for GC-2030 and GC-2010
    221-41444-05
    221-41444-05
    No Standard Split/Splitless Liner for GC-2025 and GC-2014
    221-75193
    221-75193
    Yes Standard Split/Splitless Liner for GC-2010

     

    Each P/N includes 5 deactivated liners. (Standard deactivation: max. temp. 400 °C)

    All liners presented here are compatible with all Shimadzu SPL-models. However, we recommend using the corresponding liner for each injector, as the position of the wool (or the indentations where the wool would be) has been specifically designed for each injector. 

     
     
    Liners with fixed wool

    The wool remains in the correct position, which is critical for into wool injections.
     

    Specification & Ordering Information

    P/N   Wool Description
    221-75188
    221-75188
    Yes Split/Splitless Liner with fixed wool - “onto wool”
    221-75189
    221-75189
    Yes Split/Splitless Liner with fixed wool - “into wool”
    221-75187
    221-75187
    Yes Tapered Split/Splitless Liner with fixed wool - "onto wool"
    221-75191
    221-75191
    Yes Tapered Split/Splitless Liner with fixed wool - "into wool"


    Each P/N includes 5 deactivated liners. Suitable for all injector models. (Standard deactivation: max. temp. 400 °C)